Target SCX (II)
Target SCX (II) special column uses strong cation exchange bonded silica gel as filler. The filler is optimized on the basis of original SCX. The uniform spherical particles have double functions of ion exchange and hydrophobicity. In the separation process under Pharmacopoeia chromatographic conditions, the peak symmetry of arecoline is good, and the theoretical plate number reaches 12282, which is higher than the theoretical plate number specified in Pharmacopoeia. According to the calculation of arecoline peak, it should not be lower than 3000.
Conditions of areca nut chromatographic column
Column:Target scx(II) 5 µm, 4.6 x 250 mm
Mobile phase:Acetonitrile : 0.2% Phosphoricacid ( adjust pH to 3.8 by Ammonia ) = 55 : 45
Flow rate:0.6 mL/min
Temperature:Room temperature
Detection:UV 215 nm
Injectionvolume:10 µL
Sample:0.1 mg/mL Arecoline hydrobromide in mobile phase
name |
retention time |
the measure of area |
Tailing factor |
Number of trays |
Arecoline |
7.83 |
5308 |
0.99 |
12282 |
Part No |
Product description |
Sample test |
301805-4625 |
Target -scx(II) 5 µm, 4.6 x 250 mm |
Areca,Leonurus,guanidine dicarboxylate, etc |