Pharmacopoeia column

Target -Ephedrine

      The special column for ephedra was filled with polar ether and phenyl bonded silica gel, and the packing was uniform spherical particle with carbon loading of 11.0%. In theseparation process under Pharmacopoeia chromatographic conditions, the peak symmetry of ephedrine and pseudoephedrine is good, and the number of theoretical plates is more than 10000, which is far higher than the requirement of "the number of theoretical plates should not be less than 3000 calculated according to the peak of ephedrine hydrochloride" stipulated in Pharmacopoeia.


Chromatographic conditions of ephedrine hydrochloride
Column: Target - ephedrine (5um, 4.6 × 250 mm)
Mobile phase: 0.092% phosphoric acid solution (containing 0.04% triethylamine and 0.02% di-n-butylamine-methanol (98.5:1.5) (V / V)
Flow rate: 1.0ml/min
Column temperature: RT
Detection: UV 210 nm
Injection volume: 10ul
Sample
 1. Ephedrine hydrochloride (0.03768 mg / ml)
 2. Pseudoephedrine hydrochloride (0.04648 mg / ml)



name

retention time

Area

Resolution

Tailing factor

Number of trays

unknown

3.75

4.04

 

1.79

3837.7

ephedrine

9.39

43.16

18.52

0.93

10649.1

Pseudoephedrine

11.56

52.79

5.18

1.29

10144.2

 

Part No

Product description

Sample test

301803-4625

Target -Ephedrine 5um  250*4.6mm

Special column for Ephedra


* If you need other specifications that are not in this list, please contact your sales representative.
Contact Feedback
Tel:06021-58830-11
Sitemap( Google
Copyright:MS Technologies Tel:06021-58830-0  Fax:06021-58830-11    Add:Zeppelinstr. 5, 63741 Aschaffenburg, Germany

MS Technologies